Light Measurement for Holography全息术中的光测量
One of the important parameters for transmission holography is the ratio between the light intensities of the reference beam and the object beam. It has been found that having the reference beam brighter by the ratio 4:1 to 10:1 works well for holography. You can perhaps visualize the reason for a stronger reference beam by saying that you need a strong carrier beam and a smaller modulating beam from the object to produce the image. 传输全息术的一个重要参数是参考光束与物体光束光强的比值。发现参考光束比物体光束亮4:1到10:1时,对于全息术效果较好。你或许可以这样理解更强的参考光束:你需要一个强的载波光束和一个较小的物体光束来产生图像。
For our holography experiments, it is important to optimize exposure for the film used. The film emulsion was Agfa 8E75HD which has optimum exposure at 25 microjoules/cm2. The Thorlabs detector shown above is rated at 0.47 amps/watt at 640 nm, so this sensitivity was used in calculating exposures. Since the detector area is 0.01 cm2, the exposure time in seconds is calculated from 在我们的全息实验中,优化曝光对于所用的胶片至关重要。该胶片乳剂为Agfa 8E75HD,其最佳曝光量为25微焦/平方厘米。上图所示的Thorlabs探测器在640纳米处的灵敏度为0.47安/瓦,因此在计算曝光量时使用了这一灵敏度。由于探测器面积为0.01平方厘米,曝光时间(秒)是根据以下公式计算的:
![]() When we measured the light directly from the reference beam at the film position, and then added the light from the object, we found that we got better exposures by increasing the exposure 2 stops. That is, we multiplied the calculated time by four. 当我们直接从参考光束在胶片位置测得的光,再加上物体的光,我们发现通过将曝光时间增加2档(即四倍)可以获得更好的曝光效果。也就是说,我们将计算出的时间乘以了四。
|
Index Holography concepts 索引全息概念 | ||||
| HyperPhysics***** Class Home ***** Physics 3901
HyperPhysics ***** 课程主页 ***** 物理学3901 |
Go Back
返回 |